Avoiding rotational mismatch of locking distal tibia plates depends on proper plate position
Hyung-Keun Song1, Jin-Wook Noh, Jae-Hoo Lee
1Department of Orthopaedic Surgery, Ajou University School of Medicine, Suwon, Korea.
Objectives:
The purpose of this study was to evaluate the most appropriate positioning of anatomical locking plates on the distal tibia.
Methods:
We positioned 4 types of locking plates (3 straight and 1 bayonet) at anterior, middle, and posterior positions on the medial malleolus on both lower limbs of 15 cadavers. We measured the rotational mismatch angle between the tibia and the proximal part of the plate and calculated the plate prominence at each position.
Results:
The average rotational mismatch angles of a combined group containing all 3 straight plates were 0.3, 16.9, and 29.4 degrees at the anterior, middle, and posterior positions, respectively (P < 0.001). In the group with 1 bayonet plate, the rotational mismatch angles were -10.5, 0, and 16.7 degrees on average, respectively (P < 0.001). The average plate prominence for each of the positions (anterior, middle, and posterior) in the group with 3 straight plates were 3.5, 6.9, and 9.4 mm, respectively (P < 0.001). The average plate prominence in the group with 1 bayonet plate were 5.4, 3.6, and 6.7 mm, respectively (P < 0.001).
Conclusions:
We confirmed that anterior placement of the 3 straight locking plates in the distal tibia showed the best fit with regard to rotational alignment and that middle placement of the bayonet locking plates showed best fit. Placement of distal end of the locking plates in inappropriate positions results in a greater risk of rotational mismatch and plate impingement.
Related Concept Videos
Bones of the Lower Limb: Femur and Patella
Bones of the Lower Limb: Tibia and Fibula
Screw: Problem Solving
To evaluate the...
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...