Downstream hepatic arterial blood pressure changes caused by deployment of the surefire antireflux expandable tip
Steven C Rose1, Steven G Kikolski, James E Chomas
1Department of Radiology 8756, UCSD Medical Center, University of California, San Diego Health Sciences, 200 West Arbor Drive, San Diego, CA, 92103-8756, USA, scrose@ucsd.edu.
Purpose:
The purpose of this work was to evaluate blood pressure changes caused by deployment of the Surefire antireflux expandable tip. The pressure measurements are relevant because they imply changes in hepatoenteric arterial blood flow within this liver compartment during hepatic artery delivery of cytotoxic agents.
Methods:
After positioning the Surefire antireflux system in the targeted hepatic artery, blood pressure was obtained initially with the tip collapsed (or through a femoral artery sheath), then again after the tip was expanded before chemoembolization or yttrium 90 ((90)Y) radioembolization.
Results:
Eighteen patients with liver malignancy underwent 29 procedures in 29 hepatic arteries (3 common hepatic, 22 lobar, 4 segmental). Systolic, diastolic, and mean blood pressure were all decreased by a mean of 29 mm Hg (p = 0.000004), 14 mm Hg (p = 0.0000004), and 22 mm Hg (p = 0.00000001), respectively.
Conclusion:
When the Surefire expandable tip is deployed to prevent retrograde reflux of agents, it also results in a significant decrease in blood pressure in the antegrade distribution, potentially resulting in hepatopedal blood flow in vessels that are difficult to embolize, such as the supraduodenal arteries.
Related Concept Videos
Portal Hypertension
Measurement of Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...