Related Experiment Videos
Source of the slippage in the universal joints of the Hoffmann external fixator
F L Drijber1, J B Finlay, T K Moroz
1Orthopaedic Research Laboratory, University Hospital, University of Western Ontario, London, Canada.
Abstract:
An investigation was conducted to determine what improvements in the resistance to slippage could be obtained in selected interfaces (rod/clip torsional, cheek/bowl and cheek/clip) of the Hoffmann external fixator. The modification involved changing the standard wing-nut clamp for a bolt with a thread of 7 mm and a 1 mm pitch and placing an FAG 28-303 thrust-bearing (needle roller and cage assembly) between the bolt and the cheek. The results showed a significant improvement in the slippage values of all interfaces; increases of approximately six times were obtained at all torque values of the wing-nut clamp or fastener tested. Such improvements would markedly increase the reliability of external fixation systems and thus reduce the incidence of loss-of-reduction of fracture due to slippage of the universal joint.