Precision of pulmonary vein diameter measurements assessed by CE-MRA and steady-state-free precession imaging

Michael Groth1, Peter Bannas, Marc Regier

  • 1Department of Diagnostic and Interventional Radiology, Center for Radiology and Endoscopy, University Medical Center Hamburg-Eppendorf, Martinistr. 52, 20246 Hamburg, Germany. groth.michael@googlemail.com

European Radiology
|December 21, 2012
PubMed
Abstract

Insights

Unenhanced magnetic resonance imaging (MRI) and contrast-enhanced MRI provide comparable pulmonary vein diameter measurements. However, both methods show significant intra- and interobserver variability, requiring careful clinical interpretation.

Area of Science:

  • Radiology
  • Cardiovascular Imaging
  • Medical Physics

Background:

  • Pulmonary vein (PV) diameter is crucial for planning and follow-up of atrial fibrillation ablation.
  • Accurate PV diameter assessment is essential for effective treatment strategies.

Purpose of the Study:

  • To evaluate the inter- and intraobserver reliability of PV diameter measurements.
  • To compare contrast-enhanced magnetic resonance angiography (CE-MRA) with ECG-gated 2D multislice unenhanced steady-state-free precession (multislice SSFP) sequences.

Main Methods:

  • Sixty PV diameters in 17 patients were measured using both CE-MRA and multislice SSFP.
  • Measurements were performed in transverse and coronal orientations by two independent observers.
  • Inter- and intraobserver reliability were assessed using Bland-Altman analysis and F-test statistics.

Main Results:

  • Intraobserver limits of agreement ranged from ±0.40 cm to ±0.86 cm for both methods.
  • Interobserver agreement showed limits of agreement between ±0.34 cm and ±0.83 cm.
  • No significant differences in intra- or interobserver variances were found between CE-MRA and multislice SSFP (p > 0.05).

Conclusions:

  • Multislice SSFP and CE-MRA offer comparable precision for PV diameter measurements.
  • Both techniques exhibit considerable intra- and interobserver variability.
  • PV diameter measurements by MRI require careful consideration in clinical practice.