Dose reduction in oncological staging multidetector CT: effect of iterative reconstruction
M Karpitschka1, D Augart, H-C Becker
1Department of Clinical Radiology, Ludwig-Maximilians-University Munich, Munich, Germany. Martina.Karpitschka@med.uni-muenchen.de
Objective:
To compare radiation exposure and image quality of oncological staging multidetector CT (MDCT) examinations of the chest, abdomen and pelvis with and without iterative reconstruction (IR).
Methods:
40 patients with known malignancy underwent staging CT examinations at two time points. Both CT scans were performed on the same scanner (SOMATOM® Definition Flash, Siemens Healthcare, Forchheim, Germany). For the baseline scan, the tube current-time product was set to 250 mAs [image reconstruction: filtered back projection (FBP)] and for the follow-up scan to 150 mAs [reconstruction: iterative reconstruction (IR)]. Effective radiation doses were estimated based on dose-length products for both baseline and follow-up. Noise measurements in defined regions were compared for FBP and IR. Images were also subjectively evaluated for image quality by three radiologists with different levels of experience.
Results:
Dose reduction was 44.4±8.2% for reduced-dose CT scans with IR compared with baseline with FBP. Image noise was not significantly different between images reconstructed with FBP and IR. The subjective quality of standard-dose FBP images and reduced-dose iteratively reconstructed CT images were identical.
Conclusion:
Our results show the dose-reducing potential of IR of CT image data in oncological patients.
Advances In Knowledge:
The algorithm tested in the present scientific study allows a >45% dose reduction at maintained image quality.