Blockage of a mechanical aortic valve leaflet with Bioglue: a case report
1Department of Cardiovascular Surgery, University Medical Center Ljubljana, Ljubljana, Slovenia. tone.gabrijelcic@mf.uni-lj.si
Abstract:
Mechanical aortic valve dysfunction is a very rare event and is usually due to thrombosis, pannus overgrowth, or both. BioGlue as a cause for such a complication has been reported only occasionally. We describe a case of a 63-year-old woman who underwent operation for symptomatic tight aortic stenosis. After implantation of an aortic valve (AGN-751, size 19; St. Jude Medical, St. Paul, MN, USA) because of a transverse tear of the aortic wall above the annulus occurring during the suturing of the aortotomy, a triangular Vascutek Dacron patch (Vascutek/Terumo, Inchinnan, Scotland, UK) was included. To secure hemostasis, BioGlue (CryoLife, Kennesaw, GA, USA) was applied. A transthoracic echocardiography (TTE) examination performed after signs of ischemia appeared in the electrocardiogram on postoperative day 5 revealed an aortic transvalvular gradient of 74/38 mm Hg and a functional valve area of 1.0 cm2. No coronary lesions were revealed in a coronarography evaluation, but cinefluoroscopy (CF) examination revealed immobility of 1 valve leaflet. The reoperation revealed a thick, rough layer of the glue on the inner side of the patch. This glue had run down to the valve, blocking a mechanical leaflet. Cleaning the valve was not possible, and the valve had to be changed. The subsequent postoperative course was uneventful. The transvalvular gradient was 39/20 mm Hg, and the functional valve area was 1.2 cm2. We believe that the use of BioGlue and other surgical sealants is justified to secure complex suture lines and for maintaining hemostasis in cardiac surgery, but some precautionary rules must be respected. Authors have indicated that the glue enters through the needle holes in such cases, but our findings suggest it can also pass to the Dacron patch itself. CF is superior to TTE and transesophageal echocardiography for analyzing movement of the mechanical valve leaflet, and cardiac catheterization is rarely needed.
Related Concept Videos
Mitral Stenosis III: Medical Management
Aortic Regurgitation III: Medical Management
Aortic Regurgitation I: Introduction
Aneurysm III: Interprofessional Care
Mitral Valve Prolapse I: Introduction
Aortic Regurgitation II: Clinical Features and Diagnostic Tests