Related Experiment Video
Updated: May 15, 2026

Quantification of the Immunosuppressant Tacrolimus on Dried Blood Spots Using LC-MS/MS
Published on: November 8, 2015
Development of a liquid chromatography tandem mass spectrometry method for iothalamate measurement to assess renal
Jeanne M Rhea1, James C Ritchie, Ross J Molinaro
1Department of Pathology and Laboratory Medicine, Emory University School of Medicine, Atlanta, GA 3032, United States.
Background:
Chronic kidney disease often goes undetected due to the insensitivity of current methods to accurately assess glomerular filtration rate (GFR) in early stages of renal dysfunction. The clearance of exogenously introduced iothalamate, a commonly used radiopaque agent, is an alternative to inulin clearance for the assessment of renal function and its use in calculating GFR can serve as a screening tool for kidney transplant donors.
Methods:
A method was developed to measure iothalamate in plasma and urine samples by HPLC combined with electrospray positive ionization tandem mass spectrometry (MS/MS). Iothalamate is isolated from plasma by methanol extraction and urine using a quick-spin filtration approach, then monitored by multiple reaction monitoring using the hydrogen adduct mass transitions. Iohexol was used as an internal standard.
Results:
Iothalamate was measured within an analytical run time of 5 min, with a lower limit of quantification of 18.75 ng/ml. The intraassay and interassay variations of the plasma and urine iothalamate assays were both <9%. Recovery from plasma and urine samples ranged from 93.6% to 104.1%. GFR was calculated using the patient's urine flow rate and plasma and urine iothalamate values. Linear correlations tested by LC-MS/MS and an accepted capillary electrophoresis (CE) assay showed similar results (GFR, r=0.92, Sy/x=10.3).
Conclusions:
We developed and validated an LC-MS/MS method for quantitating iothalamate in plasma and urine to calculate GFR used for screening potential kidney donors in our hospital system. A less sensitive mass spectrometry system does not sacrifice analytical or clinical sensitivity for measuring GFR.
Related Concept Videos
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Acute Kidney Injury IV: Diagnostic Studies and Prevention
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...