Phthalate analysis by gas chromatography-mass spectrometry: blank problems related to the syringe needle
Milena Marega1, Konrad Grob, Sabrina Moret
1Department of Food Science, University of Udine, Via Sondrio 2/A, 33100 Udine, Italy. milena.marega@uniud.it
Abstract:
For the analysis of the most commonly encountered phthalates by gas chromatography-mass spectrometry (GC-MS), absorption of phthalates from the laboratory air on the outer wall of the syringe needle is shown to be an important contribution to the blank problems. It was investigated for programmed temperature vapourizing (PTV) injection. Cleaning of the needle in automated injection is of modest efficiency, since the needle cannot be immersed deeply enough into the wash vial. Two approaches were studied to minimize the transfer into the column: (i) cleaning of the needle in the injector prior to splitless injection by inserting the needle in split mode while the precolumn is backflushed, which presupposes a high injector temperature to be efficient; (ii) injection under conditions minimizing thermal desorption from the needle wall, i.e. fast injection at low injector temperature (e.g. 40 °C). Both approaches resulted in blank levels of around 0.1 pg for diisobutyl phthalate (DIBP) and dibutyl phthalate (DBP), and of around 1 pg for di(2-ethylhexyl) phthalate (DEHP). They could be useful tools in existing or future methods for the analysis of phthalates or other compounds causing blank problems through contamination of the laboratory air.
Related Concept Videos
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II