Effective patient dose during neuroradiological C-arm CT procedures
Mei Bai1, Xianghua Liu, Bin Liu
1Department of Biomedical Engineering, Xuanwu Hospital of Capital Medical University, Beijing, China. baimei12@yahoo.com
Purpose:
We aimed to determine the effective dose given to patients during neuroradiological C-arm computed tomography (CT) procedures.
Materials And Methods:
Measurements were performed on 48 patients using a dose-area product (DAP) meter. A PC-based Monte Carlo program (PCXMC, STUK, Helsinki, Finland) was used to calculate the effective dose from the DAP values of each patient. Organ doses were also measured with thermoluminescent dosimeters (TLDs) using a human-shaped phantom.
Results:
The difference between the organ doses measured using TLDs and PCXMC was not significant (P > 0.05). The mean DAP for 48 patients was 9.41±2.50 Gy·cm2; the mean effective dose for all procedures was 0.30±0.08 mSv. The coefficient for the correlation (R2) between the DAP and the effective dose was 0.97. The conversion factor between the effective dose and DAP was 0.030-0.035 mSv·Gy(-1)·cm(-2).
Conclusion:
DAP can be used as a dose indicator to calculate the organ dose and effective dose of patients based on Monte Carlo simulation. This method can provide important information on the absorbed dose and enhance the radiation protection of patients during C-arm CT procedures.
More Related Videos
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...