Introduction:
The CELL-DYN Emerald is a compact bench-top hematology analyzer that can be used for a three-part white cell differential analysis. To determine its utility for analysis of human and mouse samples, we evaluated this machine against the larger CELL-DYN Sapphire and Sysmex XT2000iV hematology analyzers.
Methods:
120 human (normal and abnormal) and 30 mouse (normal and abnormal) samples were analyzed on both the CELL-DYN Emerald and CELL-DYN Sapphire or Sysmex XT2000iV analyzers. For mouse samples, the CELL-DYN Emerald analyzer required manual recalibration based on the histogram populations.
Results:
Analysis of the CELL-DYN Emerald showed excellent precision, within accepted ranges (white cell count CV% = 2.09%; hemoglobin CV% = 1.68%; platelets CV% = 4.13%). Linearity was excellent (R² ≥ 0.99), carryover was minimal (<1%), and overall interinstrument agreement was acceptable for both human and mouse samples. Comparison between the CELL-DYN Emerald and Sapphire analyzers for human samples or Sysmex XT2000iV analyzer for mouse samples showed excellent correlation for all parameters.
Conclusion:
The CELL-DYN Emerald was generally comparable to the larger reference analyzer for both human and mouse samples. It would be suitable for use in satellite research laboratories or as a backup system in larger laboratories.