Computed tomography scan to detect intra-articular air in the knee joint: a cadaver study to define a low radiation
Sanjit R Konda1, Daniel O Howard, Soterios Gyftopoulos
1Department of Orthopaedic Surgery, NYU Hospital for Joint Diseases, New York, New York 10003, USA.
Objectives:
To determine the lowest estimated radiation dose necessary to reproducibly detect intra-articular air in the knee joint of a cadaver model.
Methods:
Ten adult fresh-frozen cadaver knees with intact joint capsules were thawed and scanned at 5 decreasing radiation doses (decreasing by approximately half from 8.42 to 0.74 mGy) after introducing increasing volumes (0, 0.1, 0.3, 0.5, 0.7, and 0.9 cc) of intra-articular air. Scans were performed using 2.0-mm slice thickness from the distal one-third of the femur to the proximal one-third of the tibia. Sagittal and coronal reconstructions of each scan using 1.0-mm slice thickness were rendered. All scans were reviewed by (1) a single attending radiologist, (2) a single attending orthopedic surgeon, and (3) a single chief resident, for the presence of intra-articular air.
Results:
The sensitivity and specificity of computed tomography scan to detect intra-articular air at each volume of intra-articular air (0.1, 0.3, 0.5, 0.7, and 0.9 cc) was 100% at 0.74 mGy--the radiation threshold dose (scan parameters: voltage 80 kV, current: 33 mA, and scan time: 12.17 seconds). The effective radiation dose at 0.74 mGy for a CT scan of the knee is approximately 0.10 mSV.
Conclusions:
Computed tomography scan to detect traumatic knee arthrotomies can be successfully accomplished at a threshold radiation dose of 0.74 mGy and for an intra-articular volume of 0.1 cc of air. This low radiation dose protocol and volume of intra-articular air should be taken into consideration with future studies evaluating the use of CT scan to detect traumatic arthrotomies.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation I: X-ray and CT