Quantification of serum free light chain kappa and lambda by the SPAPLUS analyser
Diane Maisin1, Thibault Lepoutre, Damien Gruson
1Department of Clinical Chemistry, Cliniques Universitaires Saint Luc and Université Catholique de Louvain, Brussels, Belgium. diane.maisin@uclouvain.be
Clinical Biochemistry
|January 8, 2013
Summary
The SPAPLUS® system shows good agreement with the BNII® for measuring serum free light chains (FLC) kappa and lambda. Consistent methodology is recommended for monitoring monoclonal gammopathies.
Area of Science:
- Clinical chemistry
- Immunodiagnostics
Background:
- Accurate measurement of serum free light chains (FLC) kappa (κ) and lambda (λ) is crucial for diagnosing and monitoring monoclonal gammopathies.
- The SPAPLUS® system offers a turbidimetric method for FLC determination, necessitating comparison with established immunonephelometric assays like BNII®.
Purpose of the Study:
- To clinically assess the performance of the SPAPLUS® system for serum κ FLC and λ FLC determination.
- To compare the SPAPLUS® system against the BNII® system for FLC measurements and the κ/λ ratio.
Main Methods:
- Analysis of 126 serum specimens using both the SPAPLUS® (turbidimetry) and BNII® (immunonephelometry) analysers.
- Comparison of absolute values for serum κ FLC and λ FLC, and the FLC κ/λ ratio between the two methods.
- Evaluation of results in conjunction with patient clinical history.
Main Results:
- Regression analysis revealed no significant differences between SPAPLUS® and BNII® for κ FLC and λ FLC in normal and pathological ranges.
- Minor discrepancies in absolute light chain values were observed for some patients, potentially impacting clinical interpretation.
- Overall good concordance was found between the two FLC measurement methods.
Conclusions:
- The SPAPLUS® system demonstrates good agreement with the BNII® for serum FLC analysis.
- Consistent use of the same laboratory and method is recommended for serial monitoring of patients with monoclonal gammopathies.
- FLC results require interpretation alongside other laboratory data and patient diagnosis.