Noise and signal scaling factors in digital holography in weak illumination: relationship with shot noise
M Lesaffre1, N Verrier, M Gross
1Institut Langevin: Unité Mixte de Recherche (UMR) 7587 Centre National de la Recherche Scientifique (CNRS), Ecole Supérieure de Physique et de Chimie Industrielle de Paris (ESPCI), ParisTech, 1 rue Jussieu, Paris 75005, France.
Abstract:
We have performed off-axis heterodyne holography with very weak illumination by recording holograms of the object with and without object illumination in the same acquisition run. We have experimentally studied how the reconstructed image signal (with illumination) and noise background (without) scale with the holographic acquisition and reconstruction parameters that are the number of frames and the number of pixels of the reconstruction spatial filter. The first parameter is related to the frequency bandwidth of detection in time, the second one to the bandwidth in space. The signal to background ratio varies roughly like the inverse of the bandwidth in time and space. We have also compared the noise background with the theoretical shot-noise background calculated by Monte Carlo simulation. The experimental and Monte Carlo noise background agree very well with each other.
Related Concept Videos
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Upsampling
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are slanted or...
Aliasing
If the sampling frequency is below the Nyquist rate, these replicas overlap, preventing the original signal...
Downsampling
The Fourier transform of the decimated sequence reveals a combination of scaled and shifted versions of the original spectrum. This...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...