Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Automated control of end-tidal inhalation anaesthetic concentration using the GE Aisys Carestation™
1Tom Bryson Department of Anaesthesia, Liverpool Women's Hospital, Liverpool L8 7SS, UK.
Background:
Automated control of end-tidal inhalation anaesthetic concentration is now possible. The EtControl™ module of an Aisys Carestation Anaesthetic machine digitally adjusts fresh gas flow and plenum vaporizer output to achieve a target end-tidal concentration.
Methods:
We evaluated EtControl in clinical practice by measuring volatile agent consumption and the need for user input. We compared these values with contemporaneous controls using manual control of fresh gas flow rates.
Results:
A total of 321 patients were anaesthetized with EtControl and 168 with manual control of fresh gas flow. The mean [95% confidence interval (CI)] sevoflurane usage for cases of 20-40 min duration was 14 (13-16) ml h(-1) with EtControl and 30 (26-35) ml h(-1) with manual control. For cases of the same duration, the mean (95% CI) desflurane consumption was 27 (21-33) ml h(-1) with EtControl and 45 (29-62) ml h(-1) with manual control. The average number of keypresses per case was 6.5 with EtControl and 13.6 during manual control of fresh gas flow.
Conclusions:
Automatic implementation of low-flow anaesthesia using EtControl allows the user to set and maintain a desired end-tidal volatile concentration while using less volatile agent.
More Related Videos
Related Concept Videos
Inhalational Anesthetics: Overview
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...