Reduced-dose low-voltage chest CT angiography with Sinogram-affirmed iterative reconstruction versus standard-dose
François Pontana1, Julien Pagniez, Alain Duhamel
1Departments of Thoracic Imaging and Medical Statistics, Hôpital Calmette (EA 2694 Université Lille Nord de France, Boulevard Jules Leclercq, 59037 Lille, France.
Purpose:
To evaluate image quality of low-voltage chest computed tomographic (CT) angiography with raw data-based iterative reconstruction (sonogram-affirmed iterative reconstruction) in comparison with image quality of standard-dose standard-voltage filtered back projection (FBP) CT.
Materials And Methods:
This prospective study was approved by the institutional review board, and the informed consent requirement was waived. Eighty consecutive patients who were referred for follow-up chest CT angiography underwent reduced-dose CT (hereafter, T2 examination) under technical conditions similar to those of the initial examination (hereafter, T1 examination), except the voltage selection was reduced by 20 kV with adaptation of the tube current to ensure a 50% reduction in CT dose index, and regular FBP was replaced by iterative reconstruction with sonogram-affirmed iterative reconstruction. The two techniques were compared by using paired tests (Student t test, Wilcoxon test, or McNemar test, according to the nature of variables).
Results:
When compared with standard-dose T1 studies, reduced-dose T2 images showed: (a) significantly less objective noise at the level of the trachea on mediastinal and lung parenchymal images (P < .001) and no significant difference in objective noise at the level of the aorta on mediastinal images (P = .507); (b) significantly higher signal-to-noise and contrast-to-noise (P < .001) ratios; (c) similar visual perception of noise on mediastinal (P = .132) and lung (P = .366) images, mainly rated as moderate; and (d) similar overall subjective image quality (P = .405).
Conclusion:
Raw data-based iterative reconstruction yielded equivalent subjective and improved objective image quality of low-voltage half-dose CT angiograms compared with standard-dose FBP CT images for an average dose-length product of less than 80 mGy · cm in this population.
Supplemental Material:
http://radiology.rsna.org/lookup/suppl/doi:10.1148/radiol.12120414/-/DC1.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...