Related Experiment Video
Updated: May 15, 2026

Rat Model of Right-Sided Cardiac Remodeling and Arrhythmia Using Pulmonary Artery Banding
Published on: August 30, 2024
The optimal lead insertion depth for esophageal ECG recordings with respect to atrial signal quality
Andreas Haeberlin1, Thomas Niederhauser, Thanks Marisa
1ARTORG Center for Biomedical Engineering Research, University of Bern, Bern, Switzerland.
Background:
Diagnosing supraventricular arrhythmias by conventional long-term ECG can be cumbersome because of poor p-waves. Esophageal long-term electrocardiography (eECG) has an excellent sensitivity for atrial signals and may overcome this limitation. However, the optimal lead insertion depth (OLID) is not known.
Methods:
We registered eECGs at different lead insertion depths in 27 patients and analyzed 199,716 atrial complexes with respect to signal amplitude and slope. Correlation and regression analyses were used to find a criterion for OLID.
Results:
Atrial signal amplitudes and slopes significantly depend on lead insertion depth. OLID correlates with body height (rSpearman=0.71) and can be estimated by OLID [cm]=0.25*body height[cm]-7cm. At this insertion depth, we recorded the largest esophageal atrial signal amplitudes (1.27±0.86mV), which were much larger compared to conventional surface lead II (0.19±0.10mV, p<0.0001).
Conclusion:
The OLID depends on body height and can be calculated by a simple regression formula.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...