A new tonometer--the Corvis ST tonometer: clinical comparison with noncontact and Goldmann applanation tonometers
Jiaxu Hong1, Jianjiang Xu, Anji Wei
1Department of Ophthalmology, Eye, Ear, Nose, and Throat Hospital, School of Shanghai Medicine, Fudan University, Shanghai, China.
Purpose:
To compare intraocular pressure (IOP) measurements obtained using the Topocon noncontact tonometer (NCT), the Goldmann applanation tonometer (GAT), and the Corvis ST (CST), a newly developed tonometer with features of visualization and measurement of the corneal deformation response to an air impulse. A secondary objective was to assess the agreement among the devices.
Methods:
Fifty-nine participants, including glaucoma patients (36 cases) and control volunteers (23 cases), were enrolled. One eye was selected randomly for further study. IOP measurements were obtained with the CST, NCT, and GAT by two experienced clinicians. IOP values were compared. Intraobserver variability and interobserver variability were assessed by the coefficient of variation and intraclass correlation coefficient. Device agreement was calculated by Bland-Altman analysis.
Results:
Mean IOP for all examined eyes was 18.9 ± 5.8 mm Hg for CST, 21.3 ± 6.8 mm Hg for NCT, and 20.3 ± 5.7 mm Hg for GAT. There was no statistically significant difference in IOP measurements among the tonometers except between the CST and NCT. Correlation analysis showed a high correlation between each pair of devices (all P < 0.001). The CST displayed the best intraobserver variability and interobserver variability. Bland-Altman analysis revealed a bias between CST and GAT, CST and NCT, and GAT and NCT of -1.3, -2.4, and -1.1 mm Hg, with 95% limits of agreement of -6.2 to 3.5 mm Hg, -10.1 to 5.2 mm Hg, and -8.3 to 6.2 mm Hg, respectively.
Conclusions:
The CST offers an alternative method for measuring IOP. IOP measurements taken with these devices may not be interchangeable.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Glaucoma: Overview