Enzyme entrapped polypyrrole modified electrode for flow-injection determination of glucose
M Trojanowicz1, W Matuszewski, M Podsiadła
1Department of Chemistry, University of Warsaw, Poland.
Biosensors & Bioelectronics
|January 1, 1990
Abstract:
Immobilization of glucose oxidase in electropolymerized polypyrrole film on the surface of a platinum wire electrode, provides a convenient sensor for flow-injection glucose determination. An upper limit of linear response for 100 microliters injected sample volume was estimated as 20 mM, whereas a 500 microliters injected sample volume gave an estimated detection limit of 0.5 mM. A simple electrode preparation procedure allows quick electrode renewal before each series of measurements.