Repeatability and reproducibility of corneal thickness using SOCT Copernicus HR
Silvia Vidal1, Valentín Viqueira, David Mas
1Department de Optica, Universidad de Alicante, San Vicente del Raspeig, Alicante, Spain.
Corneal thickness measurements using SOCT Copernicus HR are highly repeatable and reproducible, showing excellent agreement between observers and methods. However, epithelial measurements were less reliable, and the device has ergonomic limitations for anterior segment scanning.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Medical Imaging
Background:
- Accurate corneal thickness measurement is crucial for diagnosing and managing various ocular conditions.
- Fourier domain Optical Coherence Tomography (OCT) offers advanced imaging capabilities for anterior segment analysis.
Purpose of the Study:
- To evaluate the reliability of corneal thickness measurements obtained with the SOCT Copernicus HR device.
- To assess the intra-observer repeatability and inter-observer reproducibility of the SOCT Copernicus HR for pachymetry.
Main Methods:
- Corneal thickness (central and paracentral) and epithelial thickness were measured in 30 healthy eyes using SOCT Copernicus HR.
- Intra-class correlation coefficient (ICC) and Bland-Altman analysis were employed to assess repeatability, reproducibility, and agreement between manual and automatic measurements.
Main Results:
- High intra-observer agreement (ICC up to 0.99) and inter-observer reproducibility (ICC up to 0.99) were observed for central corneal thickness.
- Agreement between manual and automatic measurement methods was also high (ICC > 0.91).
- Repeatability and reproducibility for epithelial thickness measurements were found to be poor.
Conclusions:
- SOCT Copernicus HR demonstrates high repeatability and reproducibility for corneal pachymetric mapping.
- The device's ergonomic design may require adjustments for optimal anterior segment scanning.
- Epithelial thickness measurements using this device require further investigation due to poor reliability.