Errors in measurements of stroke volume for invasive and echo-Doppler evaluations of valvular regurgitant fractions.

S Iliceto1, G D'Ambrosio, A Amico

  • 1Division of Cardiology, University of Bari, Italy.

Insights

This study compared cardiac catheterization (CATH) and 2D echo-Doppler (ED) for measuring stroke volumes (SV) to assess valvular regurgitation. Both methods showed systematic errors, with ED and CATH differences of 11.9 +/- 16.7 ml and 19.6 +/- 20.1 ml, respectively.

Area of Science:

  • Cardiology
  • Medical Imaging
  • Diagnostic Techniques

Background:

  • Accurate stroke volume (SV) measurement is crucial for calculating valvular regurgitant fractions.
  • Cardiac catheterization (CATH) and 2D echo-Doppler (ED) are common methods, but their comparative accuracy for SV is debated.
  • A 'gold standard' for SV measurement in this context is lacking, necessitating studies on error analysis.

Purpose of the Study:

  • To compare the accuracy of CATH and ED in determining stroke volumes (SV) for calculating aortic and mitral regurgitant fractions.
  • To estimate the systematic and random errors associated with both CATH and ED methods for SV measurement.
  • To evaluate the reproducibility of SV measurements using computer simulations based on known variabilities.

Main Methods:

  • Comparison of SV measurements between CATH (thermodilution, angiography) and ED (aortic/mitral annulus, time velocity integrals) in patients without valvular regurgitation.
  • Calculation of the mean difference and standard deviation between SV measurements from the two methods to quantify errors.
  • Computer simulations using 50,000 hypothetical patients to assess the contribution of measurement reproducibility to random errors in CATH and ED.

Main Results:

  • A mean difference of 11.9 +/- 16.7 ml was observed for ED, while CATH showed a difference of 19.6 +/- 20.1 ml.
  • Significant systematic errors were identified in both ED and CATH methods.
  • Computer simulations indicated specific variability percentages for different CATH and ED measurement components.

Conclusions:

  • Both CATH and ED exhibit systematic errors when measuring stroke volumes for regurgitant fraction calculations.
  • The study quantifies the differences and errors inherent in each method, aiding in clinical interpretation.
  • Understanding these errors is vital for accurate echocardiographic and invasive hemodynamic assessments.

Related Concept Videos

Ultra-High-Speed Western Blot using Immunoreaction Enhancing Technology05:59

Ultra-High-Speed Western Blot using Immunoreaction Enhancing Technology

An ultra-high-speed western blotting technique is developed by improving the kinetics of antigen-antibody binding through cyclic draining and replenishing (CDR) technology in conjunction with an immunoreaction enhancing...
6.4K
Quantification of Circulating Pig-Specific DNA in the Blood of a Xenotransplantation Model07:34

Quantification of Circulating Pig-Specific DNA in the Blood of a Xenotransplantation Model

In this protocol, porcine specific primers were designed, plasmids-containing porcine specific DNA fragments were constructed, and standard curves for quantitation were established. Using species-specific primers, cpsDNA was quantified by qPCR in pig-to-mouse cell transplantation models and pig-to-monkey artery patch transplantation...
4.4K
Abbiategrasso Brain Bank Protocol for Collecting, Processing and Characterizing Aging Brains12:28

Abbiategrasso Brain Bank Protocol for Collecting, Processing and Characterizing Aging Brains

This protocol describes a method to track individual brain-aging trajectories through a brain donation program and proper characterization of brains. Brain donors are involved in a long-term longitudinal study including serial multi-dimensional assessments. The protocol contains a detailed description of brain processing and an accurate diagnostic...
18.1K
Rapid Fractionation and Isolation of Whole Blood Components in Samples Obtained from a Community-based Setting11:31

Rapid Fractionation and Isolation of Whole Blood Components in Samples Obtained from a Community-based Setting

We outline a methodology for the processing of whole blood to obtain a variety of components for further analysis. We have optimized a streamlined protocol that enables rapid, high-throughput simultaneous processing of whole blood samples in a non-clinical...
16.5K
Dried Blood Spot Collection of Health Biomarkers to Maximize Participation in Population Studies07:20

Dried Blood Spot Collection of Health Biomarkers to Maximize Participation in Population Studies

Biomarkers are directly-measured biological indicators of disease or health. In population and social sciences, biomarkers need to be easy to obtain, transport, and analyze. Dried Blood Spot (DBS) collection meets this need, can be collected in the field with high response rates and analyzed for a variety of...
37.1K
A Practical and Novel Method to Extract Genomic DNA from Blood Collection Kits for Plasma Protein Preservation10:43

A Practical and Novel Method to Extract Genomic DNA from Blood Collection Kits for Plasma Protein Preservation

We are describing a new method of isolating genomic DNA from whole blood collected for plasma/serology. After plasma collection, the compacted blood is usually discarded. Our novel method represents a significant improvement over existing methods and makes DNA and plasma available from a single collection, without requesting additional...
42.7K