Related Experiment Video
Updated: May 15, 2026

Employing Digital Droplet PCR to Detect BRAF V600E Mutations in Formalin-fixed Paraffin-embedded Reference Standard Cell Lines
Published on: October 8, 2015
Sensitive electrochemical analysis of BRAF V600E mutation based on an amplification-refractory mutation system
1Department of Laboratory Medicine, Nanfang Hospital, Southern Medical University, Guangzhou 510515, PR China.
Abstract:
A novel electrochemical biosensor was developed for the analysis of BRAF V600E mutation in colorectal cancer cell samples based on a dual amplification strategy of amplification-refractory mutation system (ARMS) PCR and multiple enzyme labels. The labeled amplicons were conjugated on Fe3O4/Au nanoparticles using Au-S linkages. Alkaline phosphatases were then loaded onto the nanoparticles through biotin-streptavidin interactions. The resultant composite nanoparticles were characterized by transmission electron microscopy, electrochemical impedance spectroscopy, and cyclic voltammetry. In the presence of 2-phospho-l-ascorbic acid, the mutant alleles were quantified on a screen-printed carbon electrode (SPCE) from the anodic current of the enzymatic product, ascorbic acid. BRAF V600E mutant alleles concentrations as low as 0.8% were successfully determined in an excess of wild-type background. In a cell-line dilution model, the proposed method was more sensitive than were DNA sequencing and agarose gel electrophoresis. This work demonstrates a new strategy for sensitivly detecting BRAF V600E variations. It can pave the way for analyzing other rare mutations in complex cancer samples because of its high sensitivity, simplicity, low cost, and easy validation of assay procedures.