Determination of renal stone composition in phantom and patients using single-source dual-energy computed tomography
Naveen M Kulkarni1, Brian H Eisner, Daniella F Pinho
1Abdominal Imaging and Intervention, Harvard Medical School, Massachusetts General Hospital, Boston, MA 02114, USA. .
Purpose:
This study aimed to characterize the urinary tract stones in phantom and patients using single-source dual-energy computed tomography.
Materials And Methods:
Twenty stones of pure crystalline composition (uric acid [UA], struvite, cystine, and calcium oxalate monohydrate) were assessed in a phantom and 11 patients (age 39-67 years) with urinary tract stones were evaluated. An initial low-dose unenhanced CT (tube potential, 120 kilovolts [peak]; milliampere range, 150-450; noise index, 26; section thickness, 5 mm) followed by a targeted dual-energy computed tomography acquisition on a single-source dual-energy computed tomography (Discovery CT 750 HDCT, GE) was performed. Uric acid and non-UA stones were defined using a 2-material decomposition (material density-iodine/water) algorithm. The stone effective atomic number (Zeff) was used to subclassify non-UA stones. The stone attenuation (Hounsfield unit) was also studied to determine its performance in predicting the composition. Ex vivo chemical analysis of the stone served as a criterion standard.
Results:
Of the 59 verified stones (phantom, 20; patients, 39; mean size, 6 mm), there were 16 UA and 43 non-UA type. The material density images were 100% sensitive and accurate in detecting UA and non-UA stones. The Zeff accurately stratified struvite, cystine, and calcium (calcium oxalate monohydrate) stones in the phantom. In patients, Zeff identified 83% of calcium stones (n = 24), and in stones of mixed type, it resembled dominant composition. The Hounsfield unit measurements alone were 71% sensitive and 69% accurate in detecting the UA stones.
Conclusions:
Single-source dual-energy computed tomography can accurately predict UA and non-UA stone composition in vitro and in vivo. Substratification of non-UA stones of pure composition can be made in vitro and in vivo. In stones of mixed composition, the Zeff values reflect the dominant composition.
Related Concept Videos
Imaging Studies III: Computed Tomography
Urinary Tract Calculi III: Medical Management
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Urinary Tract Calculi II: Pathophysiology and Clinical Manifestations
Imaging Studies II: Ultrasonography