Related Experiment Video
Updated: May 15, 2026

Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Disparity between mainstream and sidestream end-tidal carbon dioxide values and arterial carbon dioxide levels
Murat Pekdemir1, Orhan Cinar, Serkan Yilmaz
1Department of Emergency Medicine, Faculty of Medicine, Kocaeli University, Kocaeli, Turkey. mpekdemir@yahoo.com
Background:
Measuring and monitoring end-tidal carbon dioxide (PETCO2) is an important aspect of caring for critically ill patients. The 2 methods used for PETCO2 measurement are the mainstream and sidestream methods.
Objective:
To assess the agreement between PETCO2 measurements performed by mainstream and sidestream methods with the PaCO2 values.
Methods:
This was a prospective observational study. A total of 114 subjects were enrolled in the study. PETCO2 measurements using mainstream and sidestream methods were performed simultaneously with the arterial blood sampling in subjects who were observed in the emergency department and required arterial blood gas analysis. Agreement between the PETCO2 measurements and the PaCO2 values obtained from arterial blood gas analysis were evaluated using the Bland-Altman method.
Results:
Sixty subjects (52.6%) were female, and the mean age was 60.9 years (95% CI 58.3-63.6). The mean PaCO2 was 35.16 mm Hg (95% CI 33.81-36.51), the mainstream PETCO2 was 22.11 (95% CI 21.05-23.18), and the sidestream PETCO2 was 25.48 (95% CI 24.22-26.75). Bland-Altman analysis showed an average difference between mainstream PETCO2 and PaCO2 values of 13 mm Hg (95% limits of agreement -0.6 to 25.5) and moderate correlation (r = 0.55, P < .001). The average difference between the sidestream PETCO2 and PaCO2 values was 9.7 mm Hg (95% limits of agreement -5.4 to 24.7) and poor correlation (r = 0.41, P < .001).
Conclusions:
PETCO2 values obtained by mainstream and sidestream methods were found to be significantly lower than the PaCO2 values. There was essentially no agreement between the measurements obtained by 2 different methods and the PaCO2 values.
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and HCO3− values are examined to...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Carbon Dioxide Transport in the Blood
Forms of CO2 Transport
1. Dissolved in plasma: A small percentage (7-10%) of CO2 is transported and dissolved directly in the plasma.
2. Carbaminohemoglobin: Just over 20% of CO2 is chemically bound to...
Disorders of Acid-Base Balance
Respiratory Acidosis and Alkalosis
Respiratory acidosis occurs due to an increase in the partial pressure of carbon dioxide PCO2 in the blood. It often arises from shallow breathing or impaired gas exchange caused by...
Acute Respiratory Failure-III