A study of CT monochromatic imaging for quantitative detecting hemoglobin levels
Related Experiment Video
Updated: May 15, 2026

A Rapid and Chemical-free Hemoglobin Assay with Photothermal Angular Light Scattering
Published on: December 7, 2016
Xingwang Wu1, Wanqin Wang, Le Wang
1Department of Radiology, The First Affiliated Hospital of Anhui Medical University, Hefei, Anhui, China.
Purpose:
The purpose of this study was to investigate the value of spectral CT monoenergetic imaging for detecting hemoglobin levels.
Material And Methods:
Sixty-five hospitalized patients received chest non-contrast CT scan in gemstone spectral imaging (GSI) mode on a GE Discovery CT750 HD. This study was approved by the hospital ethics committee and informed consent was signed by every patient. Raw data were reconstructed at 1.25 mm and then transferred to a AW4.4 workstation. The CT value of the ascending aorta at 40 keV∼140 keV was measured under GSI viewer of AW4.4. Hemoglobin contents were measured biochemically within 24 hrs after CT scan. The SPSS 16.0 software package was used to analyze the correlation between the obtained CT value and hemoglobin level.
Results:
At 40 keV ∼ 100 keV, the CT value decreased followed by the increase of keV and gradually stabilized at 100 keV ∼ 140 keV with an amplitude variation of 10 HU. The CT values measured at 40 keV ∼ 140 keV were well correlated with hemoglobin levels. The optimal correlation was observed at the monoenergetic level of 70 keV (r=0.633, p=0.000). The relationship between CT value and hemoglobin content could be expressed as CT value=13.015+0.245 × hemoglobin content. At 40 keV ∼ 140 keV, there is strong linear correlation between the CT value of the ascending aorta and hemoglobin content. The optimal linear relationship was observed at 70 keV.
Conclusion:
Spectral CT monoenergetic imaging can be applied for quantitative determination of hemoglobin content within a specified area of the circulatory system.
More Related Videos
07:06Simultaneous Evaluation of Cerebral Hemodynamics and Light Scattering Properties of the In Vivo Rat Brain Using Multispectral Diffuse Reflectance Imaging
Published on: May 7, 2017
09:31Automated Quantification of Hematopoietic Cell – Stromal Cell Interactions in Histological Images of Undecalcified Bone
Published on: April 8, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...