Related Experiment Video
Updated: May 15, 2026

Enumeration of Major Peripheral Blood Leukocyte Populations for Multicenter Clinical Trials Using a Whole Blood Phenotyping Assay
Published on: September 16, 2012
Intra- and interlaboratory variability of paroxysmal nocturnal hemoglobinuria testing by flow cytometry following the
Iuri Marinov1, Martina Kohoutová, Vlasta Tkáčová
1Institute of Hematology and Blood Transfusion, Prague, Czech Republic. Iuri.Marinov@uhkt.cz
Background:
Sutherland et al. recently published the Practical Guidelines for high-sensitivity detection of paroxysmal nocturnal hemoglobinuria (PNH) clones by flow cytometry (FCM), containing concise protocols for PNH testing.
Methods:
Using this approach, we studied the intra- and interlaboratory variability observed in a multicenter study in which fresh blood samples containing three clinically relevant PNH clone sizes within the granulocytic, monocytic, and red blood cell (RBC) populations were shipped to each participating center.
Results:
Coefficients of variation (CVs) for precision/reproducibility analysis ranged from 0.01%/0.02% to 0.48%/0.45% (big clone), from 0.69%/1.52% to 4.24%/5.80% (small-intermediate clone), from 1.47%/3.91% to 15.01% /17.83% (minor clone) for PNH white blood cells (WBCs) and from 0.24%/0.48% to 1.76%/1.83% (big clone), from 0.80%/1.14% to 2.39%/4.45% (small-intermediate clone), from 1.09%/3.36% to 10.54%/10.23% (minor clone) for PNH RBCs, respectively. Linear regression analysis showed excellent performance correlation between centers (r > 0.99), Wilcoxon rank test revealed no statistically significant differences for PNH granulocytes, monocytes, and RBCs (P > 0.05%), Bland-Altman analysis demonstrated good performance agreement for all target PNH clones (mean bias ranging from -1.47 to 0.71).
Conclusion:
Our results demonstrate very good intra- and interlaboratory performance characteristics for both precision and reproducibility analyses and excellent correlation and agreement between centers for all target PNH clone sizes. Our data confirm the reliability and robustness of the recently published Practical Guidelines approach for high sensitivity PNH testing by flow cytometry and suggest that such an approach represents an excellent basis for standardization of PNH testing by flow cytometry.