[Comparison of testing methods for reticulated platelets].

Yao-ling Luo1, Jing-ping Zhang, Miao-huan Kuang

  • 1Department of Clinical Laboratory, Sun Yat-sen University Cancer Center, Guangzhou, China.

Zhonghua Yi Xue Za Zhi
|January 19, 2013
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Retraction: Shaoyao-Gancao Decoction alleviated hyperandrogenism in a letrozole-induced rat model of polycystic ovary syndrome by inhibition of NF-κB activation.

Bioscience reports·2026
Same author

Asymmetric Glycol-Ether Molecule Design Enables Temperature-Adaptive Anion Coordination for Durable High-Temperature Sodium Batteries.

Journal of the American Chemical Society·2026
Same author

Depression is a risk factor for traumatic brain injury: Evidence from the National Health and Nutrition Examination Survey and Mendelian randomization analysis.

Chinese journal of traumatology = Zhonghua chuang shang za zhi·2026
Same author

A Bio‑Chemical Synergy Strategy: Mycelium Separator Breaks the Coupled Degradation Cycle in Aqueous Zn‑I<sub>2</sub> Batteries.

Advanced materials (Deerfield Beach, Fla.)·2026
Same author

Tailoring polyaniline with dual dopant engineering as a high efficiency cathode material for aqueous zinc ion batteries.

Journal of colloid and interface science·2025
Same author

Benchmarking spleen tyrosine kinase (Syk) expression in T- and NK-cell neoplasms: Implications for therapeutic targeting.

British journal of haematology·2025
Same journal

[Systematic review and meta-analysis of the prevalence of small intestinal bacterial overgrowth in patients with functional gastrointestinal disorders].

Zhonghua yi xue za zhi·2026
Same journal

[Dendritic cell and/or PD-1 monoclonal antibody therapy induces intratumoral remodeling of CD8⁺T cells in humanized cerebral glioblastoma mice models].

Zhonghua yi xue za zhi·2026
Same journal

[Study on expression characteristics of complement C3 in childhood asthma and its mechanism regulating macrophage migration and polarization involved in immune inflammatory response].

Zhonghua yi xue za zhi·2026
Same journal

[Consistency evaluation of the domestic Lifotronic platform and foreign luminescence platforms in detecting core cerebrospinal fluid biomarkers in Chinese Alzheimer disease patients].

Zhonghua yi xue za zhi·2026
Same journal

[Analysis of clinical characteristics and risk factors in children and adolescents with narcolepsy type 1 across different severities of daytime sleepiness].

Zhonghua yi xue za zhi·2026
Same journal

[Standard operating procedure for pre-analytical blood biomarker processing in Alzheimer's disease (2026 edition)].

Zhonghua yi xue za zhi·2026
See all related articles

The Sysmex XE-5000 analyzer offers more accurate and stable detection of reticulated platelets compared to flow cytometry. This automated method is recommended for clinical laboratories due to its reliability and ease of use.

Area of Science:

  • Hematology
  • Clinical Laboratory Science
  • Diagnostic Technology

Background:

  • Accurate enumeration of reticulated platelets is crucial for diagnosing and monitoring various hematological conditions.
  • Existing methods for reticulated platelet detection present varying levels of accuracy, stability, and potential for cross-contamination.
  • Establishing reliable local reference ranges is essential for accurate clinical interpretation.

Purpose of the Study:

  • To compare the accuracy, stability, and sample cross-contamination of two distinct methods for reticulated platelet percentage detection.
  • To establish a local normal reference range for reticulated platelets.
  • To provide methodological recommendations for clinical laboratory practice.

Main Methods:

  • Peripheral blood samples from healthy individuals were analyzed using the Sysmex XE-5000 analyzer (polymethyl oxazine staining) and a flow cytometer (thiazole orange staining).

More Related Videos

Routine Screening Method for Microparticles in Platelet Transfusions
09:49

Routine Screening Method for Microparticles in Platelet Transfusions

Published on: January 31, 2018

Microfluidic Flow Chambers Using Reconstituted Blood to Model Hemostasis and Platelet Transfusion In Vitro
10:25

Microfluidic Flow Chambers Using Reconstituted Blood to Model Hemostasis and Platelet Transfusion In Vitro

Published on: March 19, 2016

Related Experiment Videos

Last Updated: May 15, 2026

Microfluidics in Assessing Platelet Function
06:47

Microfluidics in Assessing Platelet Function

Published on: November 8, 2024

Routine Screening Method for Microparticles in Platelet Transfusions
09:49

Routine Screening Method for Microparticles in Platelet Transfusions

Published on: January 31, 2018

Microfluidic Flow Chambers Using Reconstituted Blood to Model Hemostasis and Platelet Transfusion In Vitro
10:25

Microfluidic Flow Chambers Using Reconstituted Blood to Model Hemostasis and Platelet Transfusion In Vitro

Published on: March 19, 2016

  • Spearman's nonparametric correlation was used to analyze the relationship between the two methods.
  • Stability was assessed by repeated measurements over time, and accuracy and cross-contamination were evaluated routinely.
  • Main Results:

    • The Sysmex XE-5000 exhibited a lower coefficient of variation (16.2%) compared to flow cytometry (35.1%).
    • Both methods showed similar sample cross-contamination rates (around 5%).
    • The Sysmex XE-5000 provided stable results within 4 hours post-sampling, while flow cytometry data fluctuated.
    • A significant correlation (P < 0.01, r(2) = 0.923) was observed between the methods.
    • Local normal reference ranges were established: 1.0%-7.5% for Sysmex XE-5000 and 3.0%-10.5% for flow cytometry.

    Conclusions:

    • The fully automated Sysmex XE-5000 analyzer is superior to flow cytometry for reticulated platelet detection due to its operational simplicity and data stability.
    • The Sysmex XE-5000 is recommended as the preferred method for clinical laboratories assessing reticulated platelet percentages.