Development of an artificial placenta V: 70 h veno-venous extracorporeal life support after ventilatory failure in
Brian W Gray1, Ahmed El-Sabbagh, Sara J Zakem
1Department of Surgery, University of Michigan, Ann Arbor, MI, USA. briangra@med.umich.edu
Purpose:
An artificial placenta would change the paradigm of treating extremely premature infants. We hypothesized that using a veno-venous extracorporeal life support (VV-ECLS) artificial placenta after ventilatory failure would stabilize premature lambs and maintain normal fetal physiologic parameters for 70 h.
Methods:
A near-term neonatal lamb model (130 days; term=145) was used. The right jugular vein (drainage) and umbilical vein (reinfusion) were cannulated with 10-12 Fr cannulas. Lambs were then transitioned to an infant ventilator. After respiratory failure, the endotracheal tube was filled with amniotic fluid, and VV-ECLS total artificial placenta support (TAPS) was initiated. Lambs were maintained on TAPS for 70 h.
Results:
Six of seven lambs survived for 70 h. Mean ventilation time was 57 ± 22 min. During ventilation, mean MAP was 51 ± 14 mmHg, compared to 44 ± 14 mmHg during TAPS (p=0.001). Mean pH and lactate during ventilation were 7.06 ± 0.15 and 5.7 ± 2.3 mmol/L, compared to 7.33 ± 0.07 and 2.0 ± 1.8 mmol/L during TAPS (p<0.001 for both). pO(2) and pCO(2) remained within normal fetal parameters during TAPS, and mean carotid blood flow was 25 ± 7.5 mL/kg/min. Necropsy showed a patent ductus arteriosus and no intracranial hemorrhage in all animals.
Conclusions:
The artificial placenta stabilized premature lambs after ventilatory failure and maintained fetal circulation, hemodynamic stability, gas exchange, and cerebral perfusion for 70 h.