Evaluation of vaporized hydrogen peroxide fumigation as a method for the bio-decontamination of the high efficiency
Hai Quan Jia1, Yan Ju Li, Bei Sun
1National Bio-protection Engineering Center, Tianjin 300161, China.
Objective:
To evaluate the performance of vaporized hydrogen peroxide (VHP) for the bio-decontamination of the high efficiency particulate air (HEPA) filter unit.
Methods:
Self-made or commercially available bioindicators were placed at designated locations in the HEPA filter unit under VHP fumigation. The spores on coupons were then extracted by 0.5 h submergence in eluent followed by 200- time violent knocks.
Results:
Due to the presence of HEPA filter in the box, spore recovery from coupons placed at the bottom of the filter downstream was significantly higher than that from coupons placed at the other locations. The gap of decontamination efficiency between the top and the bottom of the filter downstream became narrower with the exposure time extended. The decontamination efficiency of the bottom of the filter downstream only improved gently with the injection rate of H2O2 increased and the decontamination efficiency decreased instead when the injection rate exceeded 2.5 g/min. The commercially available bioindicators were competent to indicate the disinfection efficiency of VHP for the HEPA filter unit.
Conclusion:
This assay developed can detect all 16 β-lactams demanded by the European Union (EU). The whole procedure takes only 45 min and can detect 42 samples and the standards with duplicate analysis.