Performance of the Elecsys Rubella IgG assay in the diagnostic laboratory setting for assessment of immune status

Martin Enders1, Uwe Bartelt, Frank Knotek

  • 1Laboratory Prof. Gisela Enders and Colleagues, MVZ, Stuttgart, Germany. menders@labor-enders.de

Insights

Congenital rubella infection poses significant risks during pregnancy. The Elecsys Rubella IgG assay demonstrates high sensitivity and specificity for screening, though quantitative result variations necessitate consistent assay use.

Area of Science:

  • Immunology
  • Virology
  • Clinical Diagnostics

Background:

  • Congenital rubella syndrome (CRS) can cause severe birth defects and long-term health issues.
  • Maternal rubella immune status screening is crucial for preventing CRS.
  • Despite vaccination, rubella virus infection remains a concern in pregnancy.

Purpose of the Study:

  • To compare the performance of the Elecsys Rubella IgG assay against other routine screening methods.
  • To evaluate the sensitivity and specificity of the Elecsys assay for rubella IgG detection.
  • To assess quantitative differences in rubella IgG levels reported by various assays.

Main Methods:

  • Comparative performance evaluation of diagnostic assays for rubella IgG.
  • Testing of 1,090 antenatal care samples using Elecsys, Enzygnost, and hemagglutination inhibition (HAI) tests.
  • Additional testing of low-titer samples (<32 HAI) with Vidas, AxSYM, Liaison, and Architect assays.

Main Results:

  • All tested assays demonstrated 100.0% specificity for rubella IgG detection.
  • Elecsys, AxSYM, and Enzygnost assays showed high sensitivity (>90.0%) in samples with low HAI titers.
  • Significant quantitative variations in rubella virus IgG levels were observed between different immunoassays.

Conclusions:

  • The Elecsys Rubella IgG assay is suitable for routine automated screening due to its high sensitivity and specificity for qualitative results.
  • Considerable variation in quantitative results necessitates documenting testing methods and using the same assay throughout antenatal follow-up.
  • Consistent use of a single assay minimizes variability in rubella IgG quantification during pregnancy management.