Intrasession reproducibility of RNFL thickness measurements using SD-OCT in eyes with keratoconus
Michele Reibaldi1, Maurizio G Uva, Teresio Avitabile
1Eye Clinic, University of Catania, Catania, Italy.
Summary
Spectral-domain optical coherence tomography (SD-OCT) measurements of peripapillary retinal nerve fiber layer (RNFL) thickness are reproducible within the same session for eyes with keratoconus. This finding supports the use of SD-OCT for monitoring RNFL changes in keratoconus patients.
Area of Science:
- Ophthalmology
- Medical Imaging
- Retinal Imaging
Background:
- Keratoconus is a progressive thinning of the cornea.
- Retinal nerve fiber layer (RNFL) thickness is a key indicator of glaucoma and other optic neuropathies.
- Accurate RNFL measurements are crucial for diagnosis and monitoring.
Purpose of the Study:
- To assess the intrasession reproducibility of peripapillary RNFL thickness measurements using spectral-domain optical coherence tomography (SD-OCT).
- To compare RNFL measurement reproducibility between eyes with keratoconus and normal control eyes.
Main Methods:
- Peripapillary RNFL thickness was measured three times in one eye per participant during a single visit.
- Measurements were performed using SD-OCT with integrated eye-tracking and retest functions.
- Reproducibility was quantified using within-subject standard deviation (Sw), coefficient of variation (CV), and intraclass correlation coefficient (ICC).
Main Results:
- Overall global RNFL thickness showed high reproducibility in control eyes (Sw: 1.43 ± 0.24, CV: 1.28%, ICC: 0.969).
- Eyes with keratoconus demonstrated comparable intrasession reproducibility (Sw: 1.41-1.57, CV: 1.18-1.37%, ICC: 0.951-0.977).
- The ICC values for keratoconus eyes ranged from 0.951 to 0.977, indicating good consistency.
Conclusions:
- SD-OCT measurement of peripapillary RNFL thickness exhibits good intrasession reproducibility in eyes affected by keratoconus.
- These findings suggest SD-OCT is a reliable tool for evaluating RNFL in keratoconus patients.