The Medtronic Melody® transcatheter pulmonary valve implanted at 24-mm diameter--it works
Sharon L Cheatham1, Ralf J Holzer, Joanne L Chisolm
1The Heart Center, Nationwide Children's Hospital, Columbus, Ohio.
Objectives:
We report the Melody valve implanted and/or expanded to 24-mm diameter.
Background:
The Medtronic Melody valve has been implanted up to 22 mm in the pulmonary position for over a decade.
Methods:
A retrospective chart review was performed on 82 patients who underwent Melody valve implant. Technical implant method, pre- and postimplant echocardiographic findings, and initial follow-up were reviewed.
Results:
Between 04/2008 and 12/2011, 13 Melody valves were successfully implanted in 11 patients, median age 35 years (range 16-61 years), in the pulmonary (bioprosthetic valve, right ventricle to pulmonary artery conduit, native valve) position (n = 9), tricuspid position (bioprosthetic valve n = 3), and aortic position (bioprosthetic valve n = 1). Ten valves were delivered on a 24-mm balloon in balloon catheter and three were implanted using a 22-mm Ensemble balloon delivery system, followed by postdilation using a 24-mm × 2-cm Atlas balloon catheter. Postimplant, the median peak systolic gradient across the pulmonary valve was 7 mm Hg and median gradient across the tricuspid valve was 3 mm Hg. There was no change in gradient across the Melody valve in the aortic position where valve prosthesis-patient mismatch was present. Postimplant intracardiac echocardiography demonstrated none or mild valve regurgitation. No more than mild regurgitation was noted at a median follow-up of 9.5 months.
Conclusions:
The Melody valve can be implanted at 24 mm in the stenotic/regurgitant bioprosthetic pulmonary, tricuspid, and aortic valve, dysfunctional right ventricle to pulmonary artery conduit, and the native right ventricular outflow tract, whereas the valve remains competent with only mild regurgitation.
Related Concept Videos
Mitral Stenosis III: Medical Management
Mitral Regurgitation III: Medical Management
Mitral Stenosis I: Introduction
Mitral Valve Prolapse II: Assessment and Management
Aortic Regurgitation III: Medical Management
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...