Related Experiment Video
Updated: May 14, 2026

07:13
Fluorescent Lateral Flow Immunoassay Based on Quantum Dots Nanobeads
Published on: June 28, 2024
Lateral flow assay with near-infrared dye for multiplex detection
Christina Swanson1, Annalisa D'Andrea
1SRI International, Menlo Park, CA, USA. christina.swanson@sri.com
Clinical Chemistry
|February 1, 2013
Summary
This study introduces near-infrared (NIR) lateral flow assays (LFAs) for sensitive, quantitative biomarker detection. These multiplex NIR-LFAs show high accuracy, offering a valuable tool for point-of-care diagnostics.
Area of Science:
- Biomedical Diagnostics
- Assay Development
- Point-of-Care Testing
Background:
- Lateral flow assays (LFAs) are widely used point-of-care diagnostic tools due to their speed and ease of use.
- Traditional LFAs often suffer from limited analytical sensitivity, lack quantitative output, and face challenges in multiplexing.
- These limitations hinder their application in precise biomarker measurement.
Purpose of the Study:
- To develop a quantitative, multiplex lateral flow assay (LFA) utilizing near-infrared (NIR) detection.
- To enhance analytical sensitivity and enable simultaneous measurement of multiple biomarkers.
- To demonstrate the feasibility of NIR detection for improved LFA performance.
Main Methods:
- Conjugated NIR dye to antibodies and integrated them into LFAs for singleplex and duplex assays.
- Optimized singleplex NIR-LFAs for interleukin-6 (IL-6) detection (0–200 pg/mL).
- Developed duplex NIR-LFAs for simultaneous IL-6 (0–100 pg/mL) and C-reactive protein (CRP) (50–2500 ng/mL) measurement, validated with spiked samples against ELISA.
Main Results:
- NIR-LFAs achieved a limit of detection of 4 pg/mL for IL-6 in a plasma matrix with <7% CV.
- Duplex NIR-LFAs accurately quantified both IL-6 and CRP simultaneously.
- Assay results showed strong correlation with ELISA, with R² values of 0.9825 for IL-6 and 0.9711 for CRP.
Conclusions:
- NIR-LFAs provide quantitative measurements at picogram-per-milliliter concentrations with high signal-to-background ratios.
- The developed assays can detect biomarkers at vastly different concentrations in a multiplex format.
- NIR-LFAs represent a promising advancement for biomarker evaluation in point-of-care settings, offering comparable performance to ELISAs.