Related Experiment Video
Updated: May 14, 2026

Multimodal Cross-Device and Marker-Free Co-Registration of Preclinical Imaging Modalities
Published on: October 27, 2023
Evaluation of an image-based tracking workflow using a passive marker and resonant micro-coil fiducials for automatic
M Neumann1, E Breton, L Cuvillon
1Laboratoire des Sciences de l’Image, de l’Informatique et de la Télédétection, Strasbourg, France. m.neumann@unistra.fr
Abstract:
In this paper, an original workflow is presented for MR image plane alignment based on tracking in real-time MR images. A test device consisting of two resonant micro-coils and a passive marker is proposed for detection using image-based algorithms. Micro-coils allow for automated initialization of the object detection in dedicated low flip angle projection images; then the passive marker is tracked in clinical real-time MR images, with alternation between two oblique orthogonal image planes along the test device axis; in case the passive marker is lost in real-time images, the workflow is reinitialized. The proposed workflow was designed to minimize dedicated acquisition time to a single dedicated acquisition in the ideal case (no reinitialization required). First experiments have shown promising results for test-device tracking precision, with a mean position error of 0.79 mm and a mean orientation error of 0.24°.

