Related Experiment Video
Updated: May 14, 2026

Diffuse Reflectance Spectroscopy: Getting the Capillary Refill Test Under One's Thumb
Published on: December 2, 2017
High-throughput spectral system for interrogation of dermally-implanted luminescent sensors
1Department of Biomedical Engineering, Texas A&M University, College Station, TX 77843, USA. ruiqilong@gmail.com
Abstract:
Ratiometric luminescent microparticle sensors have been developed for sensing biochemical targets such as glucose in interstitial fluid, enabling use of dermal implants for on-demand monitoring. For these sensor systems to be deployed in vivo, a matched optoelectronic system for interrogation of dermally-implanted sensors was previously designed, constructed, and evaluated experimentally. During evaluation experiments, it revealed that the system efficiency was compromised by losses due to fiber connections of a commercial spectrometer. In this work, a high-throughput spectral system was presented to solve the photon loss problem. This system was designed, constructed, and tested. The throughput was around hundred time more than the previous system we used, and it was cost-effective, as well. It enables use of an integrated system for excitation, collection and measurement of luminescent emission, and will be used as a tool for in vivo studies with animal models or human subjects.
More Related Videos
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
10:38Monitoring Cell-autonomous Circadian Clock Rhythms of Gene Expression Using Luciferase Bioluminescence Reporters