Related Experiment Video
Updated: May 14, 2026

A Structured Rehabilitation Protocol for Improved Multifunctional Prosthetic Control: A Case Study
Published on: November 6, 2015
Preliminary functional assessment of a multigrasp myoelectric prosthesis
Skyler A Dalley1, Daniel A Bennett, Michael Goldfarb
1Department of Mechanical Engineering, Vanderbilt University, Nashville, TN 37240, USA. skyler.a.dalley@gmail.com
Abstract:
The authors have previously described a multigrasp hand prosthesis prototype, and a two-site surface EMG based multigrasp control interface for its control. In this paper, the authors present a preliminary assessment of the efficacy of the prosthesis and multigrasp controller in performing tasks requiring interaction and manipulation. The authors use as a performance measure the Southampton Hand Assessment Procedure (SHAP), which entails manipulation of various objects designed to emulate activities of daily living, and provides a set of scores that indicate level of functionality in various types of hand function. In this preliminary assessment, a single non-amputee subject performed the SHAP while wearing the multigrasp prosthesis via an able-bodied adaptor. The results from this testing are presented, and compared to recently published SHAP results obtained with commercially available single-grasp and multigrasp prosthetic hands.