Effects of the use of multiple scanners and of scanner upgrade in longitudinal voxel-based morphometry studies
Hidemasa Takao1, Naoto Hayashi, Kuni Ohtomo
1Department of Radiology, Graduate School of Medicine, University of Tokyo, Bunkyo-ku, Tokyo, Japan.
Purpose:
To evaluate the effects of inter-scanner variability (bias) and of scanner upgrade on longitudinal changes in regional gray matter volume.
Materials And Methods:
A total of 215 normal subjects were scanned twice, at an interval of approximately 1 year, using two 3.0 Tesla (T) scanners of the exact same model. Both scanners were simultaneously upgraded during the study period. The subjects were divided into four groups according to the combination of scanners used at the two time points. Using VBM, we evaluated longitudinal changes in regional gray matter volume in each group and the effects of scanner upgrade on longitudinal changes.
Results:
The use of different scanners at two time points significantly influenced longitudinal changes in regional gray matter volume. Regional gray matter volumes were relatively stable within the same scanner, but significantly different between the two scanners. Scanner upgrade had effects comparable to those of using different scanners at the two time points.
Conclusion:
The results of our study indicate that, even with scanners of the exact same model, the use of different scanners at different time points significantly influences longitudinal morphometric results, and that scanner upgrade has effects comparable to those of using different scanners at different time points.
More Related Videos
06:26Meta-analysis of Voxel-Based Neuroimaging Studies using Seed-based d Mapping with Permutation of Subject Images (SDM-PSI)
Published on: November 27, 2019
15:263D Modeling of the Lateral Ventricles and Histological Characterization of Periventricular Tissue in Humans and Mouse
Published on: May 19, 2015
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System IV: CMRI