Autophaser: an algorithm for automated generation of absorption mode spectra for FT-ICR MS
David P A Kilgour1, Rebecca Wills, Yulin Qi
1Department of Chemistry, University of Warwick, Coventry, CV4 7AL, United Kingdom.
Abstract:
Phase correction of Fourier transform ion cyclotron resonance (FT-ICR) mass spectrometry data allows the spectra to be presented in absorption mode. Absorption mode spectra offer superior mass resolving power (up to a factor of 2), mass accuracy, and sensitivity over the conventional magnitude mode. Hitherto, the use of absorption mode in FT-ICR mass spectrometry has required either specially adapted instrumentation or a manually intensive process of phase correction or has ignored the potentially significant effects of image charge and the associated frequency shifts. Here we present an algorithm that allows spectra recorded on unadapted FT-ICR mass spectrometers to be phase corrected, their baseline deviations removed, and then an absorption mode spectrum presented in an automated manner that requires little user interaction.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Tandem Mass Spectrometry
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Mass Spectrometers
Atomic Emission Spectroscopy: Lab