The BladderScan BVI 6200® is not accurate enough for use in a bladder retraining program
Goedele M Beckers1, Henricus J R van der Horst, Jos Frantzen
1Department of Urology, VU University Medical Center, De Boelelaan 1117, 1081 HV Amsterdam, The Netherlands.
The BladderScan BVI 6200 device is not accurate enough for measuring bladder volumes in children. It cannot reliably replace conventional ultrasound in bladder retraining programs.
Area of Science:
- Pediatric Urology
- Medical Device Technology
- Diagnostic Imaging
Background:
- Bladder scans are crucial for estimating urinary volumes in adult urology.
- Smaller bladder scan devices exist for pediatric use, but literature on their accuracy is conflicting.
- The BladderScan BVI 6200 was developed to improve accuracy in pediatric bladder volume measurement.
Purpose of the Study:
- To evaluate the accuracy of the BladderScan BVI 6200 in measuring pediatric bladder volumes.
- To determine if the BVI 6200 can replace conventional ultrasound in bladder retraining programs for children.
Main Methods:
- 84 children (mean age 7.8 years) participated in the study.
- Bladder volumes were measured using the BVI 6200, conventional ultrasound, and a graduated cylinder.
- Data were analyzed using Bland-Altman analysis to compare measurements.
Main Results:
- While mean volume differences were within 10%, the spread of measurements was very large.
- Significant over- and under-estimations were observed across different measured volumes.
- The device demonstrated considerable variability in accuracy.
Conclusions:
- The BladderScan BVI 6200 is not sufficiently reliable for accurate bladder volume measurement in children.
- The device is not recommended as a replacement for conventional ultrasound in pediatric bladder retraining programs.
Related Concept Videos
Imaging Studies II: Ultrasonography
Imaging Studies VI: Voiding Cystourethrography and Cystography
Urinary Bladder
In males, the bladder is situated in front of the rectum, while in females, it is positioned anterior to the vagina and uterus. The bladder floor contains an inverted triangular area called the trigone, defined by the two ureteric...
Anatomy of the Genitourinary System II: Bladder and Urethra
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Urodynamic Studies: Uroflowmetry