Limits of detection of explosives as determined with IMS and field asymmetric IMS vapour detectors
Anna Zalewska1, Wojciech Pawłowski, Waldemar Tomaszewski
1Faculty of Chemistry, Warsaw University of Technology, 3 Noakowskiego, 00-664, Warsaw, Poland.
Abstract:
Performance of two handheld trace explosives detectors types: SABRE 4000 and MO-2M was compared in terms of the limits of detection determined for: trinitrotoluene, hexahydro-1,3,5-trinitro-1,3,5-triazine, pentaerythritol tetranitrate and nitroglycerin in the particle detection mode. SABRE 4000 detector is a conventional detector based on the ion mobility spectrometry (IMS) and radioactive nickel as the ionization source, whereas in the MO-2M type detector a variant of this technique, viz. the field asymmetric ion mobility spectrometry (FAIMS), and a different β emitter, tritium, as an ionization source are employed. In determining limits of detection the sampling technique was suggested which resorts to binomial distribution for the determination of the indispensable number of positive measurements required to regard a given amount of explosives detected as reliable. Lower limits of detection were achieved in the study for trinitrotoluene and hexahydro-1,3,5-trinitro-1,3,5-triazine with the MO-2M detector (10- and 100-fold, respectively) than with the SABRE detector, whereas for the remaining two substances limits of detection were identical for either detector. The results obtained were discussed in the aspect of the effect of the detector design, ionization mechanism and volatility of substance as the factors that determine the limits of detection values.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
07:22Standardized Method for Measuring Collection Efficiency from Wipe-sampling of Trace Explosives
Published on: April 10, 2017
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Emission Spectroscopy: Lab
High-Performance Liquid Chromatography: Types of Detectors