Multimodal Volumetric Retinal Imaging by Oblique Scanning Laser Ophthalmoscopy (oSLO) and Optical Coherence Tomography (OCT)
Published on: August 4, 2018
Phase-sensitive swept-source optical coherence tomography imaging of the human retina with a vertical cavity
WooJhon Choi1, Benjamin Potsaid, Vijaysekhar Jayaraman
1Department of Electrical Engineering and Computer Science, and Research Laboratory of Electronics, Massachusetts Institute of Technology, Cambridge, Massachusetts 02139, USA.
Abstract:
Despite the challenges in achieving high phase stability, Doppler swept-source/Fourier-domain optical coherence tomography (OCT) has advantages of less fringe washout and faster imaging speeds compared to spectral/Fourier-domain detection. This Letter demonstrates swept-source OCT with a vertical cavity surface-emitting laser light source at 400 kHz sweep rate for phase-sensitive Doppler imaging, measuring pulsatile total retinal blood flow with high sensitivity and phase stability. A robust, simple, and computationally efficient phase stabilization approach for phase-sensitive swept-source imaging is also presented.