Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
Published on: December 18, 2016
B1+-mapping with the transient phase of unbalanced steady-state free precession
Carl Ganter1, Marcus Settles, Isabel Dregely
1Department of Radiology, Klinikum rechts der Isar, Technische Universität München, Munich, Germany.
Purpose:
A novel B1+-mapping technique (B1-TRAP) is presented, which derives the actual flip angle from the frequency of signal oscillations, observed in the transient phase of unbalanced steady-state free precession sequences.
Theory:
For short repetition times (TR), the angular frequency of distinct oscillations in the transient phase of steady-state free precession sequences is proven to be approximately proportional to the actual flip angle: ω⋅TR≈α. The result is not influenced by off-resonance and it can be shown that deviations are only of second order in the small parameter TR/T2.
Methods:
B1-TRAP makes use of this effect through a frequency analysis of the transient phase of a train of steady-state free precession signals.
Results:
In terms of reliability and time efficiency, a two-dimensional multislice implementation was found to be optimal. Unlike many steady-state B1+-mapping methods, the accuracy of B1-TRAP was not impaired by imperfect slice profiles.
Conclusion:
Simulations, phantom, and in vivo measurements showed that B1-TRAP offers a good compromise with respect to speed, robustness, and accuracy.
More Related Videos
Related Concept Videos
Atomic Nuclei: Larmor Precession Frequency
Atomic Nuclei: Nuclear Relaxation Processes
Atomic Nuclei: Magnetic Resonance
Interpreting ¹H NMR Signal Splitting: The (n + 1) Rule
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences