A new high-performance liquid chromatography method accurately measures cyclosporin A (immunosuppressive drug) in transplant patients' blood. This stable assay provides reliable results for therapeutic drug monitoring, improving patient care.
Area of Science:
- Clinical Chemistry
- Pharmacology
- Analytical Chemistry
Background:
- Transplant patients require precise monitoring of cyclosporin A levels due to its narrow therapeutic range.
- Existing methods may lack the stability, accuracy, or reproducibility needed for routine therapeutic drug monitoring.
Purpose of the Study:
- To develop and validate a stable, accurate, and reproducible high-performance liquid chromatographic (HPLC) method for cyclosporin A analysis in whole blood.
- To establish a reliable assay for routine monitoring of cyclosporin A in transplant recipients.
Main Methods:
- Development of an HPLC method utilizing the Varian Advanced Automated Sample Processor.
- Analysis of cyclosporin A and internal standard recovery from 200 microliters of whole blood.
- Validation of assay linearity, detection limit, and coefficient of variation.
Main Results:
- Achieved 81% absolute recovery for cyclosporin A and internal standard.
- Established a detection limit of 12.5 ng/ml and perfect linearity (r² = 1.0) over the 0-1000 ng/ml range.
- Demonstrated high precision with a coefficient of variation of 1.87% at 250 ng/ml and a cycle time of 10.2 min per sample.
Conclusions:
- The developed HPLC method is stable, accurate, and reproducible for cyclosporin A quantification in whole blood.
- The assay's efficiency allows for high-throughput analysis, processing up to eighty samples per day with results within 16 hours.
- This method provides a reliable tool for therapeutic drug monitoring of cyclosporin A in transplant patients.