Performance of infrared ear and forehead thermometers: a comparative study in 205 febrile and afebrile children
Patricia A Hamilton1, Lorenzo S Marcos, Michelle Secic
1ClinStat Consulting LLC, 2492 Manchester Ave., Cardiff by the Sea, CA 92007, USA. ph@clinstatconsulting.com
Aims And Objectives:
This study compared readings from two professional-grade, commercially available infrared (IR) thermometers, the ThermoScan® PRO 4000 prewarmed tip ear thermometer and the Temporal Scanner(™) TAT-5000 temporal artery thermometer.
Background:
The repeatability and precision of readings from IR thermometers for professional use were questioned in the past, but in recent years, these types of thermometers have been technologically improved, so their ability to replicate standard temperature readings reliably should be re-examined.
Design:
Febrile and afebrile children were recruited from the emergency department, overflow treatment areas and the paediatric intensive care unit of a large hospital in Argentina. Each child had a randomised sequence of seven temperature readings, including three from the ear, three from the forehead or behind the ear and one reference oral or rectal reading.
Methods:
Temperature readings were taken with the ThermoScan PRO 4000, the Temporal Scanner TAT-5000 and the monitor mode of SureTemp® Plus, a widely used professional-grade contact thermometer, for reference.
Results:
Of 205 children, 46% were febrile, per reference thermometer readings. While mean ThermoScan PRO 4000 febrile measurements did not differ significantly from reference, mean Temporal Scanner TAT-5000 febrile measurements were significantly lower (by a mean of 0·42 °C) than the reference. Overall bias of the ThermoScan PRO 4000 was significantly lower than that of the Temporal Scanner TAT-5000; repeatability was 1·5 times higher, and overall false-negative rate was about a third that of Temporal Scanner TAT-5000, when compared to the reference.
Conclusions:
This study indicates that the ThermoScan PRO 4000 provides measurements closer to those of a professional-grade contact thermometer when compared to the Temporal Scanner TAT-5000.
Relevance To Clinical Practice:
The odds of misclassifying a febrile child as non-febrile were about threefold higher with the Temporal Scanner TAT-5000 temporal artery thermometer than with the ThermoScan® PRO 4000 ear thermometer.
More Related Videos
14:05Behavioral Assessment of Hearing in 2 to 4 Year-old Children: A Two-interval, Observer-based Procedure Using Conditioned Play-based Responses
Published on: January 23, 2017
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
Published on: June 2, 2023
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...