Monitoring and compensating phase imperfections in cine balanced steady-state free precession
Rudolf Fritz Fischer1, Christoph Barmet, Markus Rudin
1Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland.
Purpose:
To analyze and correct for eddy current-induced phase imperfections in cardiac cine balanced steady-state free precession (bSSFP) imaging.
Methods:
Eddy current-induced phase offsets were measured for different phase-encoding schemes using a higher order dynamic field camera. Based on these measurements, offset phases were corrected for in postprocessing and by run-time phase compensation applying radiofrequency phase increments and additional compensatory gradient areas. The findings were validated using numerical simulations, phantom experiments, and in vivo cardiac scans.
Results:
Depending on the phase-encoding scheme, significant eddy current-induced phase offsets were detected. Time-varying phase offsets were observed at subsequent excitations leading to steady-state distortions and hence to profile-dependent amplitude modulations in k-space. Taking into account measured k-space trajectories algebraic image reconstruction allowed compensating imperfect spatial encoding. Correction of amplitude modulations was successfully accomplished by run-time phase compensation.
Conclusion:
Using magnetic field monitoring, artifacts in cine balanced steady-state free precession caused by uncompensated eddy current fields can be significantly reduced.
Related Concept Videos
Gyroscope: Precession
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...
Time-Domain Interpretation of PD Control
Consider the example of control of motor torque. Initially, a positive...
NMR Spectrometers: Resolution and Error Correction
Load-frequency control
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...