Related Experiment Video
Updated: May 14, 2026

Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
Published on: December 18, 2016
Bilateral kidney sodium-MRI: Enabling accurate quantification of renal sodium concentration through a two-element
Raffi Kalayciyan1, Friedrich Wetterling, Sabine Neudecker
1Computer Assisted Clinical Medicine, Heidelberg University, Mannheim, Germany. raffi.kalayciyan@medma.uni-heidelberg.de
Purpose:
To develop a sodium-MRI ((23) Na-MRI) method for bilateral renal sodium concentration (RSC) measurements in rat kidneys at 9.4 Tesla (T).
Materials And Methods:
To simultaneously achieve high B1 -field homogeneity and high receive sensitivity a dual resonator system composed of a double-tuned linearly polarized (1) H/(23) Na volume resonator and a newly developed two-element (23) Na receive array was used. In conjunction with three-dimensional (3D) ultra-short Time-to-Echo sequence a quantification accuracy of ± 10% was achieved for a nominal spatial resolution of (1 × 1 × 4) mm(3) in 10 min acquisition time. The technique was applied to study the RSC in six kidneys before and after furosemide-induced diuresis.
Results:
The loop diuretic agent induced an increase of cortical RSC by 22% from 86 ± 16 mM to 105 ± 18 mM (P = 0.02), whereas the RSC in the inner medulla decreased by 38% from 213 ± 24 mM to 132 ± 25 mM (P = 0.8×10(-4) ). The RSC changes measured in this study agreed well with the qualitative sodium signal intensity variations reported elsewhere.
Conclusion:
Furosemide-induced diuresis has been investigated accurately with herein presented quantitative (23) Na-MRI technique. In the future, RSC quantification could allow for defining pathological and nonpathological RSC ranges to assess sodium concentration changes, e.g., induced by drugs.
Related Concept Videos
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies II: Ultrasonography
Magnetic Resonance Imaging