A method was developed for the determination of (129)I in soil samples that uses an ICP-MS equipped with an octopole reaction system. Oxygen was used as the reaction gas for reducing the background intensity of m/z 129, principally by (129)Xe(+). The contribution of polyatomic ions, such as (127)IH(2)(+), could be effectively corrected for by assuming a production ratio ((127)IH(2)(+)/(127)I(+) = 3 × 10(-8)). The measured (129)I/(127)I ratios in a NIST standard solution and Fukushima soil samples are consistent with the expected value within the analytical error. This method provides a powerful tool for investigating radioiodine contamination in Fukushima and elsewhere.