Outer-Boundary Assisted Segmentation and Quantification of Trabecular Bones by an Imagej Plugin
Published on: March 14, 2018
Cone-beam computed tomography and microtomography for alveolar bone measurements
Nathália Ferrare1, André Ferreira Leite, Hugo César Pinto Marques Caracas
1Department of Dentistry, Faculty of Health Science, University of Brasília, Campus Universitário Darcy Ribeiro Asa Norte, Brasília, DF 70910-900, Brazil.
Purpose:
To compare cone-beam computed tomography (CBCT) and microtomography (micro-CT) for alveolar bone measurements.
Methods:
Forty teeth and alveolar bone blocks of five pigs were scanned on a micro-CT with a 9.05 μm pixel size, and on a CBCT device at 0.125 mm voxel size. One height and four thickness measurements were performed twice in standardized slices by two radiologists to verify reliability. Agreement between imaging methods was assessed by correlation coefficients, Bland-Altman plots, and the difference was tested by a Wilcoxon signed-rank test.
Results:
Regarding intra- and interobserver agreements, all bone measurements presented excellent precision values for micro-CT, but interobserver agreement for CBCT presented good to moderate values. Bone height differed about 0.3 mm, but no statistically significant differences were found for the bone thickness measurements.
Conclusion:
CBCT underestimated bone height. No statistically significant differences were found for bone thickness. Regions of thin bone tissue may not be visualized on CBCT images. There are risks of underestimating bone measurements with CBCT and assuming bone loss that does not exist clinically. Although the difference of the bone height measurement was small, the clinical relevance must be analyzed on how to interpret CBCT.
More Related Videos
08:39Longitudinal Micro-Computed Tomography Image Analysis for User-Defined Region of Interest in Critical-Sized Bone Defects
Published on: June 24, 2025
13:16Computed Tomography and Optical Imaging of Osteogenesis-angiogenesis Coupling to Assess Integration of Cranial Bone Autografts and Allografts
Published on: December 22, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...