Method to reduce non-specific tissue heating of small animals in solenoid coils
Ananda Kumar1, Anilchandra Attaluri, Rajiv Mallipudi
1Lambda Z Technologies, Baltimore, MD, USA.
Purpose:
Solenoid coils that generate time-varying or alternating magnetic fields (AMFs) are used in biomedical devices for research, imaging and therapy. Interactions of AMF and tissue produce eddy currents that deposit power within tissue, thus limiting effectiveness and safety. We aim to develop methods that minimise excess heating of mice exposed to AMFs for cancer therapy experiments.
Materials And Methods:
Numerical and experimental data were obtained to characterise thermal management properties of water using a continuous, custom water jacket in a four-turn simple solenoid. Theoretical data were obtained with method-of-moments (MoM) numerical field calculations and finite element method (FEM) thermal simulations. Experimental data were obtained from gel phantoms and mice exposed to AMFs having amplitude >50 kA/m and frequency of 160 kHz.
Results:
Water has a high specific heat and thermal conductivity, is diamagnetic, polar, and nearly transparent to magnetic fields. We report at least a two-fold reduction of temperature increase from gel phantom and animal models when a continuous layer of circulating water was placed between the sample and solenoid, compared with no water. Thermal simulations indicate the superior efficiency in thermal management by the developed continuous single chamber cooling system over a double chamber non-continuous system. Further reductions of heating were obtained by regulating water temperature and flow for active cooling.
Conclusions:
These results demonstrate the potential value of a contiguous layer of circulating water to permit sustained exposure to high intensity alternating magnetic fields at this frequency for research using small animal models exposed to AMFs.
More Related Videos
06:45In Vitro and In Vivo Delivery of Magnetic Nanoparticle Hyperthermia Using a Custom-Built Delivery System
Published on: July 2, 2020
09:01Magnetic-, Acoustic-, and Optical-Triple-Responsive Microbubbles for Magnetic Hyperthermia and Pothotothermal Combination Cancer Therapy
Published on: May 22, 2020
Related Concept Videos
Solenoids
Each turn in a solenoid can be approximated as a circular current carrying coil that generates a dipole moment. The...
Decreased Body Temperature
Magnetic Field of a Solenoid
Consider a solenoid with 100 turns wrapped around a cylinder of...