Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Accuracy of transcutaneous carbon dioxide monitoring in hypotensive patients
Jung-Youn Kim1, Young-Hoon Yoon, Sung-Woo Lee
1Department of Emergency Medicine, Korea University College of Medicine, , Seoul, South Korea.
Objectives:
Continuous blood gas monitoring is frequently necessary in critically ill patients. Our aim was to assess the accuracy of transcutaneous CO₂ tension (PtcCO₂) monitoring in the emergency department (ED) assessment of hypotensive patients by comparing it with the gold standard of arterial blood gas analysis (ABGA).
Methods:
All patients receiving PtcCO₂ monitoring in the ED were included. We excluded paediatric patients, patients with no ABGA results during a hypotensive event, patients whose ABGA was not performed simultaneously with PtcCO₂ monitoring, and patients who received sodium bicarbonate for resuscitation. The included patients were classified into hypotensive patients and normotensive patients. A hypotensive patient was defined as a patient showing a mean arterial pressure under 60 mm Hg. The agreement in measurement between PaCO₂ tension (PaCO₂) and PtcCO₂ were investigated in both groups.
Results:
The mean difference between PaCO₂ and PtcCO₂ was 2.1 mm Hg, and the Bland-Altman limits of agreement (bias ± 1.96 SD) ranged from -15.6 to 19.7 mm Hg in the 28 normotensive patients. The mean difference between PaCO₂ and PtcCO₂ was 1.1 mm Hg, and the Bland-Altman limits of agreement (bias ± 1.96 SD) ranged from -19.5 to 21.7 mm Hg in the 26 hypotensive patients. The weighted κ values were 0.64 in the normotensive patients and 0.60 in the hypotensive patients.
Conclusions:
PtcCO₂ monitoring showed wider limits of agreement with PaCO₂ in urgent situations in the ED environment. However, acutely developed hypotension does not affect the accuracy of PtcCO₂ monitoring.
More Related Videos
07:52Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Errors occurring during blood pressure monitoring
Several factors...
Carbon Dioxide Transport in the Blood
Forms of CO2 Transport
1. Dissolved in plasma: A small percentage (7-10%) of CO2 is transported and dissolved directly in the plasma.
2. Carbaminohemoglobin: Just over 20% of CO2 is chemically bound to...