Far-field RF powering of implantable devices: safety considerations
Rebecca A Bercich1, Daniel R Duffy, Pedro P Irazoqui
1Biomedical Engineering Department, Purdue University, West Lafayette, IN 457906, USA. rbercich@purdue.edu
Abstract:
Far-field RF powering is an attractive solution to the challenge of remotely powering devices implanted in living tissue. The purpose of this study is to characterize the peak obtainable power levels in a wireless myoelectric sensor implanted in a patient while maintaining safe local temperature and RF powering conditions. This can serve as a guide for the design of onboard electronics in related medical implants and provide motivation for more efficient power management strategies for implantable integrated circuits. Safe powering conditions and peak received power levels are established using a simplified theoretical analysis and Federal Communications Commission-established limits for radiating antennas. These conditions are subsequently affirmed and improved upon using the finite-element method and temperature modeling in bovine muscle.